PeakTech 1210 Operation Manual page 151

Digital oscilloscope with colour display
Hide thumbs Also See for 1210:
Table of Contents

Advertisement

Available languages

Available languages

Note:
1. Entering into Autoscale function and
2. At the mode of Autoscale, the oscilloscope can self-estimate "Trigger mode" (Edge, Video, and Alternate)
and "Type" (Edge, Video). If now, you press "Trigger mode" or "Type", the forbidden information will
display on the screen.
3. At the mode of XY and STOP status, pressing AUTO SET to enter into Autoscale, DSO switches to YT
mode and AUTO status.
4. At the mode of Autoscale, DSO is always in the state of DC coupling and AUTO triggering. In this case,
the forbidden information will be showing when making Triggering or Coupling settings.
5. At the mode of Autoscale, if adjust the vertical position, voltage division, trigger level or time scale of CH1
or CH2, the oscilloscope will turn off Autoscale function and if press AUTOSET again, the oscilloscope will
enter into Autoscale.
6. Turn off the submenu at the Autoscale menu, the Autoscale is off and turn on the submenu still enters into
the function.
7. When video triggering, the horizontal time scale is 50us. If one channel is showing edge signal, the other
channel is showing video one, the time scale refers to 50us as video one as standard.
8. While the Autoscale is working,below settings will be made forcibly:
(1) The unit will switch from non-main time base to main time base status.
(2) The unit will siwtch to Peak detection menu while in the state of Average sampling mode.
38. Appendix A: Technical Specifications
Unless otherwise specified, the technical specifications applied are applicable to the probe with the
attenuation switch setting 10X and the PeakTech
following two conditions at first, can these specification standards be reached.
*
This instrument should run for more than 30 minutes continuously under the specified operating
temperature.
*
If the change range of the operating temperature is up to or exceeds 5° C, open the system function menu
and execute the "Auto-calibration" procedure.
All specification standards can be fulfilled, except one(s) marked with the word "Typical".
Sampling
ACQU Mode
Sampling Rate
flicker will be on the top left corner. (flicker every 0.5 second)
®
digital oscilloscope. Only if the oscilloscope fulfill the
Sample
Peak Detection
Average
500 MS/s for each channel
150

Advertisement

Table of Contents