Commissioning
To configure an alarm, assign appropriate values to the following parameters:
Submenu: Setup → Advanced setup → Application → Alarm → Alarm 1 to 4
Parameter
Meaning / Action
Alarm mode
• Off
• On
• Latching
Alarm value source Select process variable to be checked for limit violation.
• HH alarm value
• H alarm value
Assign appropriate limit values (see diagram above).
• L alarm value
• LL alarm value
58
No alarms are generated.
An alarm disappears if the alarm condition is no longer present (taking into consideration the
hysteresis).
All alarms remain active until the user selects Clear alarm = Yes.