Directions for use
Table 3. Electromagnetic immunity (continued)
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the 790 series cordless illumination system is used exceeds the
applicable RF compliance level above, the 790 series cordless illumination system should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the 790 series cordless illumination system.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.