Directions for use
Table 2. Immunity for all EQUIPMENT and SYSTEMS
The 790 series cordless illumination system is intended for use in the electromagnetic environment
specified below. The customer or user of the 790 series cordless illumination system should assure
that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
±2 kV mains
transient (EFT)/
±1 kV I/O
burst
IEC 61000-4-4
Surge
±1 kV differential
IEC 61000-4-5
±2 kV common
Voltage dips, short
>95% dip in
interruptions, and
0.5 cycle
voltage variations
on power supply
60% dip in
input lines.
5 cycles
IEC 61000-4-11
30% dip for
25 cycles
>95% dip in
5 seconds
Power frequency
3 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
Compliance
Electromagnetic environment -
level
guidance
± 6 kV contact
Floors should be wood, concrete, or
± 8 kV air
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
±2 kV mains
Mains power quality should be that of
Not applicable
a typical commercial or hospital
environment.
±1 kV differential
Mains power quality should be that of
±2 kV common
a typical commercial or hospital
environment.
>95% dip in
Mains power quality should be that of
0.5 cycle
a typical commercial or hospital
environment. If the user of the 790
60% dip in
series cordless illumination system
5 cycles
requires continued operation during
power mains interruptions, it is
30% dip for
recommended that the 790 series
25 cycles
cordless illumination system be
powered from an uninterruptible
>95% dip in
power supply or battery.
5 seconds
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.