Specifications; Power; Mounting; Memory - Siemens SITRANS RD200 Operating Instructions Manual

Communication and displays.
Hide thumbs Also See for SITRANS RD200:
Table of Contents

Advertisement

Specifications

2.1

Power

Input voltage option 1
Input voltage option 2
Transmitter power
supply
External loop power
supply
Input impedance
Output loop resistance •
2.2

Mounting

Location
Ambient temperature
Relative humidity
Installation category
2.3

Memory

● Non-volatile
● Stores settings for minimum of ten years if power is lost
6
85 to 265 V AC, 50/60 Hz; 90 to 265 V DC, 20 W max.
UL Recognized, 5 A max, slow blow
May share one 5 A fuse among up to 6 meters
12 to 36 V DC, 12 to 24 V AC, 6 W max.
UL Recognized, 5 A max, slow blow
May share one 5 A fuse among up to 6 meters
One or two isolated trans-
mitter power supplies (op-
tional)
35 V DC maximum
Voltage ranges: greater than 1 MΩ
Current ranges: 50 - 100 Ω (depending on resettable fuse impedance)
24 V DC 10 to 700 Ω maximum
35 V DC (external) 100 to 1200 Ω maximum
Single power supply:
Dual power supplies:
Indoor/outdoor
Panel mount 1/8 DIN
Two panel mounting bracket assemblies provided
Operating temperature range: -40 to +65 °C (-40 to +149 °F)
Storage temperature range: -40 to +85 °C (-40 to +185 °F)
Relative humidity: 0 to 90% non-condensing
II
one 24 V DC ± 10% @ 200 mA maximum
one 24 V DC ± 10% @ 200 mA maximum
one 24 V DC ± 10% @ 40 mA maximum
Operating Instructions, 05/2019, A5E47629414-AA