Siemens SITRANS RD200 Operating Instructions Manual page 52

Communication and displays.
Hide thumbs Also See for SITRANS RD200:
Table of Contents

Advertisement

Setup
6.7 Advanced Features
The Internal calibration menu is part of the Advanced features menu.
1. Press and hold Right arrow
Features menu of the meter.
2. Press Up arrow
3. The meter displays either current (
setup. Press Enter/Ack
Example for current (
1. The meter displays the low input calibration (
Enter/Ack
2. After the display stops flashing, a number is displayed with the leftmost digit brighter than
the rest. The bright digit is the active digit that can be changed by pressing Up arrow
Note
For instructions on how to program numeric values see Setting numeric values
(Page 25).
3. Set the display value to correspond to the input signal being calibrated. Press Ent/Ack
Apply the high input signal in the same way that the low input signal was set in steps 1-3.
4. Press Menu
The example above shows the calibration of the current input. The voltage input is calibrated
in a similar way.
Note
• Low and high input signals can be any valid values within the range of the meter.
• Observe minimum input span requirements between input 1 and input 2.
• Low input must be less than high input signal.
52
to scroll to the Internal calibration menu (
.
Curr
) input internal calibration:
. The display flashes for a moment while meter is accepting the low input.
. Press Right arrow
to accept the display setting. The display moves to the high input calibration (
to exit at any time.
and Menu
for three seconds to access Advanced
Curr)
volt
or voltage (
to start the calibration process.
ILo
). Apply the low input signal and press
to move to the next digit.
Operating Instructions, 05/2019, A5E47629414-AA
ICAL
)and press Enter/Ack
), according to the meter input
SITRANS RD200