Connection
5.4 Current and Voltage Connections
Transmitter powered by ext. supply or self-powered
①
②
③
④
Transmitter powered by internal supply (optional)
①
②
③
④
⑤
The current input is protected against current overload by a fuse capable of being reset. The
display may or may not show a fault condition depending on the nature of the overload.
The fuse limits the current to a safe level when it detects a fault condition, and automatically
resets itself when the fault condition is removed.
16
Signal Connector
2-wire 4 to 20 mA transmitter
External power supply
2-wire 4 to 20 mA self-powered transmitter
Signal connector
24 V out
Signal
3-wire 4 to 20 mA transmitter
2-wire 4 to 20 mA transmitter
Operating Instructions, 05/2019, A5E47629414-AA
SITRANS RD200