Page 1
R e h m SIMOMED HM Service Instructions SIMOMED 44 cm / 54 cm Monitor 44 cm Monitor Part No. 37 92 079 X2077 54 cm Monitor Part No. 37 92 087 X2078 54 cm Monitor Part No. 74 11 726 X2078...
Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
When handling the monitor, there is a risk of implosion of the pic- ture tube! If not observed, death or serious bodily injury can occur. The required protective clothing must be worn. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 4 of 24 07.02...
Remarks Regarding the Monitor • The SIMOMED HM monitor recognizes the particular TV scan norm (only H scan norm) and automatically switches to it. If a norm is used that is not programmed, the monitor automatically recognizes this norm, and it may be necessary to adjust the geometry and contrast/brightness.
See CB - DOC TD00-000.801.01..(Spare Parts Catalogue) Part No. 97 16 754 Y4905 • Set of Torx offset screwdrivers. See CB - DOC TD00-000.801.01..(Spare Parts Catalogue) Part No. 99 00 663 GE999 SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 6 of 24 07.02 TD SD 21...
Page 7
See CB - DOC TD00-000.801.01..(Spare Parts Catalogue). • The Help / Adjustment texts in the service user interface can be found on the supplied service diskette under <Help>-<Service>. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 7 of 24 Medical Solutions 07.02 TD SD 21...
N/S Amplitude • A second level can be selected by pressing the key combination contrast- and the ser- vice key. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 8 of 24 07.02 TD SD 21 Medical Solutions...
Page 9
Beginning with FW 1.74 (shipped since approx. middle of ’99), keyboard blocking can be activated or deactivated by simultaneously pressing the H + and H - keys, four times in succession. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 9 of 24 Medical Solutions 07.02 TD SD 21...
NOTE For an explanation of the masks or of individual functions in a mask, a help text can be selected in the masks by clicking on a question mark button. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 10 of 24 07.02...
Insert the screw inserted in plastic as follows: Turn the screw counterclockwise until it snaps in place, only then turn it clockwise to tighten it. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 11 of 24 Medical Solutions 07.02 TD SD 21...
Configuration with Monitor Replacement Application-relevant Configurations 5Configuration with Monitor Replacement When replacing a SIMOMED HM monitor, the following configurations must be checked in the service software, relative to the application. Blocking/Enabling the Front Panel Keys In this window, all front panel keys that can be operated can be individually blocked/enabled.
(see the list - blocked Norm). Manual selection is required because the monitor cannot detect the format from the video signal. In the SSW, select under: => Configuration=> Disable Scan Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 13 of 24 Medical Solutions 07.02 TD SD 21...
Page 14
Multistar KK without Norm 1 (1023 n.a. lines/60Hz/1:1) display 2000 60Hz Multistar KK with Norm 1 (12497 Norm 12 (1249 lines/100Hz/4:3) lines/100Hz/1:1) display 2000 50Hz SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 14 of 24 07.02 TD SD 21 Medical Solutions...
Remove the rear wall of the housing Remove the 2 screws on the rear wall, take the housing off towards the back. Fig. 3: • Remove the top cover panel. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 15 of 24 Medical Solutions 07.02 TD SD 21...
Page 16
Remove the 9 screws, remove the top cover panel. Fig. 4: • Remove the rear cover panel. Remove the 7 screws, remove the rear cover panel. Fig. 5: • Jumper positions. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 16 of 24 07.02 TD SD 21 Medical Solutions...
Page 17
- 1 => ON - 2 => OFF - Standard (setting when shipped) Installation of the cover panels is in the reverse of the above instructions. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 17 of 24 Medical Solutions 07.02 TD SD 21...
• Test image - AXIOM Artis ..We use the “Siemens” test pattern (it has larger measuring fields than the SMPTE test pattern).for H/C adjustment. Select zoom to magnify the fields again. Move the corresponding field (0% - 100%) to approx.
When "Brightness / Contrast Adjustment using the front buttons or in the SSW" is performed, the ambient light sensor must be switched off in => Configuration => Power down & Ambient light sensor. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 19 of 24 Medical Solutions 07.02 TD SD 21...
Page 20
- Using the Brightness slider, set the val- ue for Max. Brightness in the 0% field from the table (Tab. 3 / p. 22). - If necessary, repeat both adjustments. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 20 of 24 07.02 TD SD 21...
Page 21
5R field and the slider for Maximum contrast to set the value in the 2L field to the values that were noted. - If necessary, repeat both adjustments. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 21 of 24 Medical Solutions 07.02 TD SD 21...
Page 22
0,5 ± 0,1 FA / FC / BA / BC 1. The monitor with Part No. 74 11 726 is not used in this application (max. luminous intensity = 450 cd/m2) SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 22 of 24 07.02...
Page 23
Changes to Previous Version 7Changes to Previous Version General Remarks Warning notes expanded. Adjustment of max. brightness / contrast Adjustment / Configuration expanded. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 23 of 24 Medical Solutions 07.02 TD SD 21...
Page 24
Changes to Previous Version SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 24 of 24 07.02 TD SD 21 Medical Solutions...