SIMOMED HM
Service Instructions
SIMOMED
44 cm / 54 cm Monitor
44 cm Monitor Part No. 37 92 079 X2077
54 cm Monitor Part No. 37 92 087 X2078
54 cm Monitor Part No. 74 11 726 X2078
Print No.:
RA52-060.061.01.07.02
Replaces: RA52-060.061.01.06.02
03792087
07411726
03792079
AX
1996
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 07.02
written

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SIMOMED HM and is the answer not in the manual?

Questions and answers

Summary of Contents for Siemens SIMOMED HM

  • Page 1 R e h m SIMOMED HM Service Instructions SIMOMED 44 cm / 54 cm Monitor 44 cm Monitor Part No. 37 92 079 X2077 54 cm Monitor Part No. 37 92 087 X2078 54 cm Monitor Part No. 74 11 726 X2078...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Adjustment ............19 7 _______ Changes to Previous Version_____________________________________ 23 Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 3 of 24 Medical Solutions 07.02...
  • Page 4: General Information

    When handling the monitor, there is a risk of implosion of the pic- ture tube! If not observed, death or serious bodily injury can occur. The required protective clothing must be worn. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 4 of 24 07.02...
  • Page 5: Product-Specific Remarks

    Remarks Regarding the Monitor • The SIMOMED HM monitor recognizes the particular TV scan norm (only H scan norm) and automatically switches to it. If a norm is used that is not programmed, the monitor automatically recognizes this norm, and it may be necessary to adjust the geometry and contrast/brightness.
  • Page 6: Troubleshooting / Repair / Replacement

    See CB - DOC TD00-000.801.01..(Spare Parts Catalogue) Part No. 97 16 754 Y4905 • Set of Torx offset screwdrivers. See CB - DOC TD00-000.801.01..(Spare Parts Catalogue) Part No. 99 00 663 GE999 SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 6 of 24 07.02 TD SD 21...
  • Page 7 See CB - DOC TD00-000.801.01..(Spare Parts Catalogue). • The Help / Adjustment texts in the service user interface can be found on the supplied service diskette under <Help>-<Service>. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 7 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 8: Front Key Panel Operation

    N/S Amplitude • A second level can be selected by pressing the key combination contrast- and the ser- vice key. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 8 of 24 07.02 TD SD 21 Medical Solutions...
  • Page 9 Beginning with FW 1.74 (shipped since approx. middle of ’99), keyboard blocking can be activated or deactivated by simultaneously pressing the H + and H - keys, four times in succession. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 9 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 10: Initial Installation Of The Service Program

    NOTE For an explanation of the masks or of individual functions in a mask, a help text can be selected in the masks by clicking on a question mark button. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 10 of 24 07.02...
  • Page 11: Fw Replacement

    Insert the screw inserted in plastic as follows: Turn the screw counterclockwise until it snaps in place, only then turn it clockwise to tighten it. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 11 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 12: Application-Relevant Configurations

    Configuration with Monitor Replacement Application-relevant Configurations 5Configuration with Monitor Replacement When replacing a SIMOMED HM monitor, the following configurations must be checked in the service software, relative to the application. Blocking/Enabling the Front Panel Keys In this window, all front panel keys that can be operated can be individually blocked/enabled.
  • Page 13: Norm Change Time

    (see the list - blocked Norm). Manual selection is required because the monitor cannot detect the format from the video signal. In the SSW, select under: => Configuration=> Disable Scan Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 13 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 14 Multistar KK without Norm 1 (1023 n.a. lines/60Hz/1:1) display 2000 60Hz Multistar KK with Norm 1 (12497 Norm 12 (1249 lines/100Hz/4:3) lines/100Hz/1:1) display 2000 50Hz SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 14 of 24 07.02 TD SD 21 Medical Solutions...
  • Page 15: Jumper Configuration (Hardware)

    Remove the rear wall of the housing Remove the 2 screws on the rear wall, take the housing off towards the back. Fig. 3: • Remove the top cover panel. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 15 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 16 Remove the 9 screws, remove the top cover panel. Fig. 4: • Remove the rear cover panel. Remove the 7 screws, remove the rear cover panel. Fig. 5: • Jumper positions. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 16 of 24 07.02 TD SD 21 Medical Solutions...
  • Page 17 - 1 => ON - 2 => OFF - Standard (setting when shipped) Installation of the cover panels is in the reverse of the above instructions. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 17 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 18: Adjusting The Video Amplifier

    • Test image - AXIOM Artis ..We use the “Siemens” test pattern (it has larger measuring fields than the SMPTE test pattern).for H/C adjustment. Select zoom to magnify the fields again. Move the corresponding field (0% - 100%) to approx.
  • Page 19: Adjustment

    When "Brightness / Contrast Adjustment using the front buttons or in the SSW" is performed, the ambient light sensor must be switched off in => Configuration => Power down & Ambient light sensor. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 19 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 20 - Using the Brightness slider, set the val- ue for Max. Brightness in the 0% field from the table (Tab. 3 / p. 22). - If necessary, repeat both adjustments. SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 20 of 24 07.02 TD SD 21...
  • Page 21 5R field and the slider for Maximum contrast to set the value in the 2L field to the values that were noted. - If necessary, repeat both adjustments. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 21 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 22 0,5 ± 0,1 FA / FC / BA / BC 1. The monitor with Part No. 74 11 726 is not used in this application (max. luminous intensity = 450 cd/m2) SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 22 of 24 07.02...
  • Page 23 Changes to Previous Version 7Changes to Previous Version General Remarks Warning notes expanded. Adjustment of max. brightness / contrast Adjustment / Configuration expanded. Siemens AG RA52-060.061.01.07.02 SIMOMED HM Page 23 of 24 Medical Solutions 07.02 TD SD 21...
  • Page 24 Changes to Previous Version SIMOMED HM RA52-060.061.01.07.02 Siemens AG Page 24 of 24 07.02 TD SD 21 Medical Solutions...