Test 1. DC Input Resistance
Procedure
1
Connect the DMM probes between the probe tip and ground at the tip of the probe.
2
Set up the DMM to measure resistance. The resistance should read 1 MW ± 3%.
3
Record the resistance in
N2797A User's Guide
Table 10
on page 46.
Performance Verification
5
39