Connecting the Chiller

Coolant Hoses and Filter Connections
The required coolant hoses, filters, and fittings are included in the plumbing kit that was
shipped with the laser. They should be connected as illustrated in Figure 2-2. The correct
coolant flow path starts with the supply port of the chillerfilter coolant in port of
laser headlaser headcoolant out port of laser head return port of the chiller.
Please be aware of the flow direction of the filter.
The filter may be attached to the back of the chiller, customer's equipment or a wall
using the provided L-bracket.
© 2019 Cutting Edge Optronics, Inc.
Figure 2-2. Filter Hoses and Coolant Connections
NOTE:
Threaded hose barbs and adapters should have threads wrapped 3-4 times with
Teflon tape.
Chapter 2: Laser System Setup
9
Patara End-Pumped Laser Quick Start Guide
Approved for Public Release; NG19-1901