First-Time Chiller Turn-On Procedure - NORTHROP GRUMMAN Patara EPL-016-QTG Quick Installation Manual

End-pumped laser
Table of Contents

Advertisement

First-time chiller turn-on procedure

The first-time turn-on procedure should be similar for use with other chillers.
Filling the Reservoir
1. Remove the filler cap from the reservoir
2.
Using a funnel, add Optishield Plus mixture (90% distilled water, 10% Optishield
Plus) until it reaches the MAX line on the reservoir's fluid level gauge. If Optishield
Plus is not allowed due to local regulations, use Optishield original formula.
3.
When full, remove the funnel, but do not replace the cap at this time.
Electrical Power
Plug the Chiller's power cord into an appropriate electrical outlet. Place the Circuit
1.
Breaker/Power Switch on the rear of the instrument enclosure to the On position.
2.
Three decimal points will appear on the Temperature display
Starting Process Fluid Flow
1.
Press the Power Button on the front panel. The system startup sequence will begin
and proceed as follows:
2.
The pump will turn on and fluid will begin circulating through the system. The set
point temperature will appear briefly on the Temperature display; after a few
seconds, it will be replaced by the actual fluid temperature.
Check for leaks
1.
Once the pump is turned on, check all of the connectors to see if there is any
leakage. If a leak is observed, turn off the pump immediately and fix the leak.
The reservoir's fluid level will drop as the process and/or process cooling lines fill
2.
with fluid. Slowly add fluid to the reservoir until the liquid level remains stable.
3.
Replace the reservoir cap
© 2019 Cutting Edge Optronics, Inc.
Chapter Three: Laser Optimization and Tuning
17
Patara End-Pumped Laser Quick Start Guide
Approved for Public Release; NG19-1901