Edrive Setup - NORTHROP GRUMMAN Patara EPL-016-QTG Quick Installation Manual

End-pumped laser
Table of Contents

Advertisement

eDrive Setup

Input Power
Use only power cords suitable for the driver. Use a power source that delivers voltage in
the range of 100 VAC (RMS) to 240 VAC (RMS) and 47 Hz to 63 Hz. Power switching
is performed automatically (i.e., there are no configuration switches to set for high or low
voltage ranging). Table 2-1 provides recommended fuse selection for each voltage range.
Mounting
The front panel of the eDrive is designed with four holes to mount into an EIA-310D-
compliant rack. If this option is used, the eDrive's weight requires extra side supports. If
the eDrive is to be used on a desktop or table top, it is recommended that the eDrive be
equipped with feet to prevent the driver from marring the surface when it is moved. The
eDrive must be secured.
Clearance
Adequate clearance should be allowed on the front, sides, and rear of the eDrive for
access to connections and components. The front and rear vents of the eDrive must be a
minimum of 24 inches (61 cm) away from walls or vertical surfaces so air flow is not
restricted.
© 2019 Cutting Edge Optronics, Inc.
Table 2-1. Recommended Fuse Ratings
AC Input
120 V / 15 A
47-63 Hz
240 V / 7 A
47-63 Hz
Frequency
Fuse Ratings
(F1, F2)
15 A
8 A
6
Patara End-Pumped Laser Quick Start Guide
Approved for Public Release; NG19-1901
Chapter 2: Laser System Setup

Advertisement

Table of Contents
loading

Table of Contents