Closed Loop Chiller - NORTHROP GRUMMAN IKA-012-QTG User Manual

Table of Contents

Advertisement

Closed Loop Chiller

The single most common cause of a laser returned for repair involves customer damage.
More than one third of these damaged lasers result from cooling problems. Coolant
problems almost always require the replacement of the diode arrays - the single most
expensive component in Northrop Grumman lasers. This section describes how to avoid
damaging arrays.
© 2019 Cutting Edge Optronics, Inc.
Approved for Public Release; NG19-2099
WARNING.
Do not operate laser without cooling. Do not power the laser diodes or Q-
switch without cooling. Inadequate heat dissipation will seriously damage the laser
diodes or Q-switch and will void warranty.
Table 3-3. Heat Capacity and Minimum Flow Rate
Model
IKA-012-QTG
IKA-016-QTG
Heat capacity should be greater than the laser's power consumption.
1
Heat Capacity
Minimum Flow Rate
1
800 W
1.5 gpm (5.7 lpm)
16
Chapter Three: System Details
Iklwa Laser User Manual