Commissioning
6.3 Adjustment and initial commissioning
Procedure
1. Measure the output resistances R
the corresponding data sheet. Example:
2. Place the test weight on all four corners: Note the measured values
3. Determine the differences from the smallest value:
1)
36
Load cells
LC 1:
LC 2:
LC 3:
LC 4:
Load cells
LC 1:
LC 2:
LC 3:
LC 4:
Load cells
Weights
Difference
LC 1: 160 kg
LC 1 - LC 2
LC 2: 148 kg
-
1)
LC 3: 155 kg
LC 3 - LC 2
LC 4: 152 kg
LC 4 - LC 2
Smallest value
of the load cells LC. Alternatively: Use the values from
o
Example values
1004.52 Ω
1003.64 Ω
1010.70 Ω
1028.12 Ω
Example values
160 kg
148 kg
155 kg
152 kg
Example values
Result
12 kg
-
7 kg
4 kg
Operating Instructions, 07/2016, A5E02199611B-03