Parameterization; Error Messages - ABB Sensyflow FMT400-VTS Commissioning Instructions

Thermal mass flowmeter
Hide thumbs Also See for Sensyflow FMT400-VTS:
Table of Contents

Advertisement

7

Parameterization

The configuration of the mass flowmeter can be changed with the aid of an LCI (local
communication interface) adapter that has to be connected to the parameterization socket.
This adapter is available as a separate accessory and comes supplied together with the
relevant software.
The following parameters can be modified or read in:
Read device data
Measurement range
Low flow
Filter factor
Output in the event of an error Minimum (lower than 3.5 mA) or maximum (greater than 22.5 mA)
Analog output
Save modified data
Factory setting
Status
Reset
Print out
Password
To activate the configuration changes, the device must be restarted. To do so, use the
command "restart reset" or briefly interrupt the power supply for approximately 10 seconds.
8

Error messages

If failures occur (e.g., sensor wire break), the 0/4 ... 20 mA output will adopt the configured
control value.
In the event of an error, the following settings can be made:
-
Minimum
-
Maximum
IMPORTANT (NOTE)
Default preset in case of a fault
Maximum
Minimum
These settings can be modified using the LC interface and the corresponding configuration
software.
Other error messages can be read out via the "status" function ("check function", section 7).
CI/FMT400-VTS/VTCS-EN
Show current settings
Selection within the calibrated measurement range
Low flow suppression 0...20 % of the upper range limit
Attenuation, 1 ... 500 can be selected (default: 50)
0 / 4 ... 20 mA
Save settings
Select default setting
Check function
Restart
Print out current settings
For service technicians only
< 3.5 mA
> 22.5 mA
0 ... 20 mA
4 ... 20 mA
Sensyflow FMT400-VTS, FMT400-VTCS
Parameterization
 
EN - 25