Table B-1: Analog voltage measurement range and resolution
1
fN1
Range
(Hz)
±5000
2000
±1000
±5000
1000
±1000
±5000
500
±1000
±5000
100
±1000
±5000
60
±1000
±5000
50
±1000
±5000
30
±1000
2
Differential measurements
(mV)
with input reversal
RMS (µV)
4.047
0.894
±200
0.310
2.838
0.613
±200
0.220
2.035
0.445
±200
0.157
0.950
0.205
±200
0.071
0.769
0.162
±200
0.056
0.732
0.161
±200
0.053
0.612
0.131
±200
0.042
Typical effective resolution
Single-ended measurements
or differential measurements
without input reversal
bits
RMS (µV)
21.3
5.806
21.2
1.290
20.4
0.459
21.9
4.359
21.7
0.931
20.9
0.343
22.3
2.999
22.2
0.659
21.4
0.248
23.4
1.388
23.3
0.323
22.5
0.131
23.7
1.140
23.6
0.261
22.9
0.113
23.8
1.112
23.7
0.254
22.9
0.111
24.1
0.901
24.0
0.217
23.3
0.099
3
bits
20.8
20.7
19.8
21.2
21.1
20.2
21.8
21.6
20.7
22.9
22.6
21.6
23.2
23.0
21.8
23.2
23.0
21.9
23.5
23.2
22.0