Specifications - Stanford Research Systems SR510 Manual

Lock-in amplifier
Table of Contents

Advertisement

General
Power
100, 120, 220, 240 VAC (50/60 Hz); 35 Watts Max
Mechanical
17" x 17" x 3.5" (Rack Mount Included) 12 lbs.
Warranty
Two years parts and labor.
Signal Channel
Inputs
Voltage:
Current:
Impedance
Voltage:
Current:
Full Scale
Voltage:
Sensitivity
Current:
Maximum
Voltage:
Inputs
Current:
Noise
Voltage:
Current:
Common Mode
Range:
Gain Accuracy
1% (2 Hz to 100KHz)
Gain Stability
200 ppm/°C
Signal Filters
60 Hz notch, -50 dB (Q=10, adjustable from 45 to 65 Hz)
120 Hz notch, -50 dB (Q=10, adjustable from 100 to 130 Hz))
Tracking bandpass set to within 1% of ref freq (Q=5)
Dynamic Reserve
20 dB LOW
40 dB NORM (100 nV to 50 mV sensitivity)
60 dB HIGH
Bandpass filter adds 20 dB to dynamic reserve
Line Notch filters increase dynamic reserve to 100 dB
Reference Channel
Frequency
0.5 Hz to 100 kHz
Input Impedance
1 MΩ, ac coupled
Trigger
SINE:
PULSE:
Mode
Fundamental (f) or 2nd Harmonic (2f)
SR510 Specification Summary
Single-ended or True Differential
6
10
Volts/Amp
100 MΩ + 25 pF, ac coupled
1 kΩ to virtual ground
100 nV (10 nV on expand) to 500 mV
100 fA to 0.5 µA
100 VDC, 10 VAC damage threshold
2 VAC peak-to-peak saturation
10 µA damage threshold
1 µA ac peak-to-peak saturation
7 nV/√Hz at 1 kHz
0.13 pA/√Hz at 1 kHz
1 Volt peak; Rejection: 100 dB dc to 1KHz
Above 1KHz the CMRR degrades by 6 dB/Octave
(1 µV to 500 mV sensitivity)
(100 nV to 5 mV sensitivity)
100 mV minimum, 1Vrms nominal
±1 Volt, 1 µsec minimum width
3