Stanford Research Systems SR510 Manual page 8

Lock-in amplifier
Table of Contents

Advertisement

Acquisition Time
25 Sec at 1 Hz
6 Sec at 10 Hz
2 Sec at 10 kHz
Slew Rate
1 decade per 10 S at 1 kHz
Phase Control
90° shifts
Fine shifts in 0.025° steps
Phase Noise
0.01° rms at 1 kHz, 100 msec, 12 dB TC
Phase Drift
0.1°/°C
Phase Error
Less than 1° above 10Hz
Demodulator
Stability
5 ppm/°C on LOW dynamic reserve
50 ppm/°C on NORM dynamic reserve
500 ppm/°C on HIGH dynamic reserve
Time Constants
Pre:
Post:
Offset
Up to 1X full scale (10X on expand)
Harmonic Rej
-55 dB (bandpass filter in)
Outputs & Interfaces
Outputs
X (RcosØ), X Offset, Noise
Output Meter
2% Precision mirrored analog meter
Output LCD
Four digit auto-ranging LCD display shows same values as the analog meters
Output BNC
±10 V output corresponds to full scale input
<1
Reference LCD
Four digit LCD display for reference phase shift or frequency
RS232
Interface controls all functions. Baud rates from 300 to 19.2 K
GPIB
Interface controls all functions. (IEEE-488 Std)
A/D
4 BNC inputs with 13 bit resolution (±10.24 V)
D/A
2 BNC outputs with 13 bit resolution (±10.24 V)
Ratio
Ratio output equals 10X output divided by the Denominator input.
Internal Oscillator
Range:
Stability:
Distortion: 2% THD
Amplitude: 1% accuracy, 500 ppm/°C stability
1msec to 100 sec (6 dB/Octave)
1sec, 0.1 sec, none (6 dB/Octave) or none
output impedance
1 Hz to 100 kHz, 1% accuracy
150 ppm/°C
4