4.5

GRAIN DETERMINING

1) Turn on the scale and wait for the scale to display zero. Press [CON/UNIT] to
select the container size you will use, an Indicator will show next to the container
size to confirm which one you have selected (for example below 0.5lt).
2) Place the empty container on the pan and press [TARE]
> 0<
Smp
stored
3) Load the 500ml container with the grain. The screen shows the net weight (the
first net weight) of the grain (in this example the total weight is 400g).
Smp
stored
At this point you can press the [CON/UNIT] key to view the yield values calculated
in k.hl and L.bu based on the net weight in the container.
4) To store this weight as the initial sample value, press the [STORE] key, the
display will show "Store smp" and then return to the last weighing mode, an
indicator above Smp/stored will show to confirm the sample weight has been
stored in the memory.
Smp
© Adam Equipment Company 2012
0.00 g
0.5lt
400.00 g
0.5lt
400.00 g
0.5lt
7
qt
Net
pt
lt
qt
Net
pt
lt
qt
Net
pt
lt