Stage I; Stage Ii; Stage Iv - Atmos ENT Operating Instructions Manual

Flexible fibre endoscope
Table of Contents

Advertisement

Stage Step
I
II
III
IV
V
Enzymatic cleaner
1
Ready-to-use solution, 0,55 % orthophtalaldehyd
2
Demineralized water (demineralised, low germs, maximum 10 germs/ml
3
and endotoxin-poor, maximum 0.25 endotoxin units/ml)
5.2.1
& The concentration of the cleaning agent depends on the contamination
of the product. Observe the manufacturer's instructions for use of the
cleaning solution.
1.
The product must be completely immersed in the cleaning solution so
that all accessible surfaces are covered.
Damaged product due to cleaning with a metal brush.
2.
Place the product in the solution and clean with a soft cloth or a
suitable soft cleaning brush.
»
There is no visible residue on the surface.
3.
Operate the movable components like focussing and defl ection lever in
the solution.
4.
Thoroughly rinse the movable components with the cleaning solution,
at least 5 times, using a disposable syringe (20 ml)
5.2.2
Use fresh water for each rinsing cycle.
18
Cleaning and disinfection
Tempera-
ture
Cleaning
35 - 45 °C /
95-113 °F
Intermedi-
Room tem-
ate rinsing
perature
(cold)
Disinfec-
20 - 25 °C /
tion
68 - 77 °F
Final
Room tem-
rinsing
perature
(cold)
Drying
Room tem-
perature

Stage I

Stage II

Time
Concen-
tration
5 min
0.8 %
3 X
---
1 min
12 min ---
3 X
---
2 min
---
---
Water
Chemistry
quality
Drinking
Cidezyme /
water
Enzol
1
Drinking
---
water
Drinking
Cidex OPA
water
VE-W
---
sterile
3
Drinking
---
water
2