Zero point adjustment
Sensor dimension
Standard Deviation Limit
DN 15
DN 25
DN 50
DN 80
DN 100
DN 150
Offset Limit
DN 15
DN 25
DN 50
DN 80
DN 100
DN 150
FC310 (From firmware 4.0)
Operating Instructions, 06/2018, A5E44036384-AA
Default value
kg/s
0,0008
0,0027
0,009
0,04
0.0832
0.1376
0,012333333
0,056166667
0,140444444
0,368611111
0.783333333
1.269444444
A.2 Sensor dimension dependent default settings
Default value
Range
kg/h
2,88
9,72
32,4
144
299.52
495.36
44,4
202,2
505,6
1327
2820
4570
Technical reference