Proservo NMS80
7.4.6
Commissioning check
This procedure is to confirm that all calibration steps have been completed appropriately.
1.
Navigate to: Diagnostics → Device check → Commissioning check → Commissioning
check
2.
Select the Start option.
Executing option is shown on the verify drum table.
3.
Select the Next option.
4.
Confirm that the Commissioning check wizard shows the Finished option.
5.
Confirm that the Verify drum table parameter is passed.
This completes the commissioning check procedure.