An annual schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized ACCU-SCOPE distributor can arrange for this service. ® ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Carefully remove the microscope from the Styrofoam container by its arm and base and place the microscope on a flat, vibration-free surface. ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
3002-CB FORENSIC COMPARISON MICROSCOPE SERIES COMPONENTS DIAGRAM Eyepiece Binocular Viewing Head Nosepiece Objective Main Switch Intensity Adjustment Knob Lock Screws Stage Main Body Power Jack Fuse Holder ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces. Eyepieces Binocular Viewing Head ④ ③ ② Objectives ① ② Main Body ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Page 7
Tighten Fig. 2 the lock screws for each as shown in Fig. 2 and Fig. 3. Fig. 3 ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Page 8
Fig. 6, and insert the pin side of the lamp completely into the lamp socket. Reinstall the collector lens housing. Collector Lens Halogen Lamp Housing Fig. 6 ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Page 9
Using a flat head screwdriver, remove the fuse holder (see Fig. 7), and unscrew the old fuse and insert the new fuse. Replace the fuse holder and retighten with the screwdriver. ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
(NOTE: the magnification of the two microscopes may change independently because of variations caused by moving the microscope or environmental changes. ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Repeat several times to check. NOTE: do not counter rotate the focusing knob as this will cause damage to the focusing system. ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Fine Focusing Coarse Focusing Use caution when changing objectives – do Knob Knob not allow the objective to touch the specimen slide. Fig. 13 ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
You will see a dark circle encroaching on the Fig. 15 bottom of the tube. ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
There are stains on the incident lens of the binocular Clean it tube The size of the aperture diaphragm is too small Adjust it The position of the condenser is too low Adjust the position ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Page 17
The binocular image is not The interpupillary distance is not correct Adjust coincident Adjust the diopter Eyes are too tired No diopter adjustment correctly ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
Page 18
Use the specified lamp to The lamp burns out Using a substandard lamp replace; if the problem is suddenly not solved, contact The voltage is too high service department ACCU-SCOPE ® 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com ...
SCOPE at (631) 864-1000. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to ACCU-SCOPE INC., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America, for all foreign warranty repairs return freight charges are the responsibility of the individual/company who returned the merchandise for repair.