Accu-Scope 3002 User Manual
Hide thumbs Also See for 3002:

Advertisement

 
 
MANUAL
 
3002
MICROSCOPE SERIES
 
73 Mall Drive, Commack, NY 11725 • 631-864-1000 (P) • 631-543-8900 (F)
www.accu-scope.com • info@accu-scope.com

Advertisement

Table of Contents
loading

Summary of Contents for Accu-Scope 3002

  • Page 1     MANUAL   3002 MICROSCOPE SERIES     73 Mall Drive, Commack, NY 11725 • 631-864-1000 (P) • 631-543-8900 (F) www.accu-scope.com • info@accu-scope.com...
  • Page 2: Table Of Contents

    EYEPIECE DIOPTER ....................16 LED LAMP SPECIFICATIONS, OPERATION, RECHARGING .......... 17 TROUBLESHOOTING LED LAMP ..................18 TECHNICAL SPECIFICATIONS ..................19 TROUBLESHOOTING ....................20-22 MAINTENANCE........................23 SERVICE ..........................23 WARRANTY ........................23 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 3: Safety Notes

    An annual schedule of preventative maintenance by qualified personnel is highly recommended. Your ® authorized ACCU-SCOPE distributor can arrange for this service. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 4: Introduction

    Carefully remove the microscope from the Styrofoam container by its arm and base and place the microscope on a flat, vibration-free surface. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 5: Components Diagram

    COMPONENTS DIAGRAM Interpupillary Distance Adjustment Iris Diaphragm Adjustment Lever Binocular Viewing Head Eyepieces Objective Coarse Focus Adjustment Knob Mechanical Stage Illuminator Housing Fine Focus Adjustment Knob On/Off Switch ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 6: Controls Diagram

    CONTROLS DIAGRAMS Interpupillary Distance Adjustment Diopter Adjustment Ring Condenser Adjustment Knob Upper Limit Stage Locking Lever Fine Focusing Knob Coarse Focusing Knob Variable Light Intensity Dial ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 7: Controls Diagram

    (continued) Y Stage Movement Adjustment Knob X Stage Movement Adjustment Knob Iris Diaphragm Adjustment Lever Fine Focusing Adjustment Knob Coarse Focusing Adjustment Knob On/Off Switch Tension Adjustment Collar ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 8: Assembly Diagram & Procedure

    When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces. Eyepieces ② Binocular Viewing Head ① ③ ④ Objectives Power Cord ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 9 Installing the Eyepieces Remove the protective caps from the eyepiece tubes. Insert the eyepieces into the eyepiece tubes as shown in Fig. 3 & 4. Fig. 3 Fig. 4 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 10 Installing the Glass Filter Swing out the condenser holder ① (Fig. 7) and insert the filter into the holder ②. Swing the holder back in place. Fig. 7 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 11 Using a flat head screwdriver, remove the fuse ⑥ holder (Fig. 9), and unscrew the old fuse and ⑦ insert the new fuse. Replace the fuse holder and retighten with the screwdriver. Fig. 10 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 12 (Fig 11), and as shown in Fig. 12, insert the pin side ② of the lamp ① completely into the lamp socket. Reinstall the illuminator housing. Fig. 12 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 13: Brightness

    Use caution when changing objectives – do not allow the objective to touch the specimen slide. Fig. 15 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 14: Focusing Tension

    Instead, rotate the right diopter collar until the sharpest image appears. Repeat several times to check. NOTE: do not counter rotate the focusing knob as this will cause damage to the focusing system. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 15: Condenser

    Fig. 18 the image of the iris diaphragm, remove one eyepiece and look through the tube. You will see a dark circle encroaching on the bottom of the tube. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 16: Interpupillary Distance

    NOTE: the diopter adjustment ring is on the LEFT eyetube. The right eyetube is fixed – DO NOT attempt to rotate the right eyetube. The diopter range is ±5. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 17: Led Lamp Specifications, Operation, Recharging

    LED Lamp Specifications The 3002-LED-R microscope is equipped with a variable 1 watt rechargeable LED lamp. The LED lamp can last up to 5 hours on a single charge depending on the brightness level used, and is rated for 10,000 hours.
  • Page 18: Troubleshooting Led Lamp

    ACCU-SCOPE dealer for service The charging circuit is not working Replace the charging circuit board call your local authorized ACCU-SCOPE dealer for service ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 19: Technical Specifications

    Eyepieces Category Magnification Field of view(mm) Φ18 Wide Field 10× Total Magnification Eyepiece 10× 10× 10× 10× Objective 4× 10× 40× 100× Total Magnification 40× 100× 400× 1000× ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 20: Troubleshooting

    Air bubbles in immersion oil Remove bubbles Condenser aperture is closed or open too much Open or close properly Position the condenser at Condenser is position too low the upper limit ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 21 Insufficient Brightness Open to the proper Aperture diaphragm closed too much setting Position the condenser at Condenser position is too low the upper limit ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 22 Tension adjustment is set too low the focusing knobs knob Loosen the tension on the Fine focus is ineffective Tension adjustment is set too high focusing knobs ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
  • Page 23: Maintenance

    SCOPE at (631) 864-1000. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to ACCU-SCOPE INC., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America, for all foreign warranty repairs return freight charges are the responsibility of the individual/company who returned the merchandise for repair.