An Allen wrench is included in the shipping carton to use for removing the screws. Failure to remove the metal plate and two screws prior to operation will cause severe damage to your mechanical stage. See diagram on next page. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
An annual schedule of preventative maintenance by qualified personnel is highly recommended. Your ® authorized ACCU-SCOPE distributor can arrange for this service. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
7. 3-prong electric power cord Optional accessories such as optional objectives and/or eyepieces, slides sets, etc., are not shipped as part of the standard equipment. These items, if ordered, are shipped separately. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
3012-LED MICROSCOPE SERIES COMPONENTS DIAGRAM 3013 Trinocular Model Digital Camera C-mount Adapter On/Off Switch Variable Light Field Iris Intensity Centering Screws Field Iris Fine Focus Knob ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
Tighten the two centering screws until the condenser is in the center of the field of view when the diaphragm is closed. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
A complete circle should be seen in the viewing field when viewing the specimen slide. An improper adjustment will cause operator fatigue and will disrupt the objective parfocality. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
Located on the left side of the stand between the coarse adjustment knob and the vertical arm is an adjustable tension control dial that is preset at our facility. This allows the user to adjust the coarse control tension to their individual preference. ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
Focus Stop The focus stop on the ACCU-SCOPE 3012-LED Series is built-in for ease of use and is pre-adjusted by our factory technicians. Lamp Replacement The LED lamp is rated for 20,000 hours and should last for many years under normal use.
Air bubbles in immersion oil Remove bubbles Condenser aperture is closed Open or close properly or open too much Condenser is positioned too Position the condenser slightly lower than the upper limit ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
Page 13
Cover glass is to thick Use a 0.17mm cover glass Diopter adjustment is not set Readjust the diopter settings properly as outlined in section 4.3 ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
Page 14
Increase the tension on the the coarse focusing knob focusing knobs Fine focus is ineffective Tension adjustment is set too Loosen the tension on the high focusing knobs ® ACCU-SCOPE 73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com...
ACCU-SCOPE INC. This warranty expressly excludes any liability by ACCU-SCOPE INC. for consequential loss or damage on any grounds, such as (but not limited to) the non-availability to the End User of the product(s) under warranty or the need to repair work processes. Should any defect in material, workmanship or electronic component occur under this warranty contact your ACCU-SCOPE distributor or ACCU-SCOPE at (631) 864-1000.