ON/OFF Control Mode Setup:
Screen # 14 - Control Action
Use the UP/DOWN arrow keys to change the value from
Heat to Cool.
Set for the particular process that is being performed.
When complete, press MENU key to advance to screen # 15.
Screen # 15 - Control Mode
Use UP/DOWN arrow keys to change the value from the
default mode of PID to On/Off.
When complete, press MENU key to advance to screen # 16.
Screen # 16 - ON/OFF Control
Use the UP/DOWN arrow keys to adjust value.
Cycle time is the rate that the output cycles on and off.
The recommended time is 1 second.
If a mechanical relay is being used from the output, it is recommended that the time be increased
to more than 1 second. Mechanical relays are slower than solid-state relays and require more
time to operate correctly. Damage to the relay and control can occur if not set correctly.
When complete, press MENU key to advance to screen # 17.
Screen # 17 - Power Output Control
Use UP/DOWN arrow keys to change the value.
Default value is 100 percent.
This feature allows the user to reduce the output power to the output heating device—useful in
applications were temperature overshoot is occurring after Auto Tune has been completed.
Experimenting with this value will be necessary because every process setup is different.
When complete, press MENU key to advance to screen # 18.