PID Control Mode Setup
Screen # 9 - Advanced Menu Gateway
Use the UP/DOWN arrow keys to change selection to ENTER.
When complete, press MENU key to advance to screen # 10.
The following screens are in the advance portion of the
setup menu:
Screen # 10 - Calibration Gateway
Screen # 11 - Global Sensor Offset
Screen # 12 - Over Temperature Stop
Screen # 13 - Loop Break Stop
Screen # 14 - Control Action
Screen # 15 - Control Mode
Screen # 16 - Auto Tune
Screen # 10 - Calibration Gateway
Screen will default to SKIP Calibration Setup.
WARNING:
you have the proper calibration equipment to calibrate the sensor inputs associated with
this controller. Changing calibration settings in the controller can cause errors in sensor
temperature readings and operation of the control. For instructions on how to use the
calibration functions, contact your instrument supplier.
With SKIP selected, press the MENU key to advance to screen # 11.
Screen # 11 - Global Sensor Offset Calibration
Use the UP/DOWN arrow keys to change the offset value.
The use of another temperature device is required to
determine the correct offset needed.
Adjusting this value is offsetting the sensor input temperature being displayed.
When complete, press MENU key to advance to screen # 12.
1.800.561.8187
:
Do not enter the calibration gateway unless
www.
Screen # 17 - Proportional Gain Setup
Screen # 18 - Integral Gain Setup
Screen # 19 - Derivative Rate Gain Setup
Screen # 20 - Power Rate Control
Screen # 21 - Run Time
Screen # 22 - Power Failure Control
12