ABB VortexMaster FSV430 Commissioning And Safety Instruction page 49

Hide thumbs Also See for VortexMaster FSV430:
Table of Contents

Advertisement

Operating mode /
Designation
(order code)
Standard volume flow
Gas Std/Norm Vol. /
(for gaseous measuring media)
NG2
Gas Mass (Density
Mass flow rate, calculated with
under reference
the density under reference
conditions) / NG3
conditions, pressure and
temperature
(for gaseous measuring media)
Gas Mass (actual
Mass flow rate, calculated using
density) / NG3
the current density in the
operating condition.
(gaseous measuring media)
2) The highest priority of the device is to record the density via the analog input, as long as the analog input is activated as a density input. If the analog input is not
available as a density input, the system attempts to record the density via the HART input. If both the analog input and the HART input are deactivated as a density
input, the system uses the default density value.
3) The connection via the analog input or HART input is described in the chapter titled "Electrical connections" on page 29.
5) The highest priority of the device is to record the temperature via the analog input, as long as the analog input is activated as a temperature input. If the analog input is
not available as a temperature input, the system attempts to record the temperature via the HART input. If both the analog input and the HART input are deactivated
as a temperature input, the system uses the default density value.
7) If for the menu item Device Setup / Plant/Customized -> Gas Std. Mode the selection is set to "Gas linear.", the compression factor is reset to 1.0. See also the
chapter titled "Special operating modes" in the operating instruction.
Additional parameters required
Operating pressure
3) 5)
Operating temperature
3) 5)
Compression factor in the standard
condition
(AGA / SGERG only)
Compression factor in the operating
condition
Reference pressure and reference
temperature in the normal condition
Density under reference conditions
Operating density
2) 3)
VortexMaster FSV430, FSV450 SwirlMaster FSS430, FSS450 | CI/FSV/FSS/430/450-EN Rev. G 49
Parameter setting
Via analog input:
Input/Output / Field Input / Analog In Value ->
Pressure
Via HART input:
Input/Output / Field Input / HART In Value ->
Pressure
Default setting for the pressure value:
Device Setup / Plant/Customized /
Compensation Setting -> Preset
Pressure(abs)
With internal temperature sensor.
No information required, the measured value
from the temperature sensor is used.
Via analog input:
Input/Output / Field Input / Analog In Value ->
Temperature
Via HART input:
Input/Output / Field Input / HART In Value ->
Temperature
Default setting for the temperature value:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Int.Temp
Adjustment via DTM/EDD
7)
Adjustment via DTM/EDD
7)
Device Setup / Plant/Customized / Gas Ref.
Conditions
Via analog input:
(For selection, see operating mode Gas
Std/Norm Vol. / NG2)
Via HART input:
(For selection, see operating mode Gas
Std/Norm Vol. / NG2)
Device Setup / Plant/Customized / Gas Ref.
Conditions, as a selection for "Ref. Density"
Via analog input:
Input/Output / Field Input / Analog In Value ->
Density
Via HART input:
Input/Output / Field Input / HART In Value ->
Density
Default setting for the density:
Device Setup / Plant/Customized /
Compensation Setting -> Preset Density