Configuration of the alarm signaling via the current output
Not for devices with Modbus-communication!
Easy Setup
Iout at Alarm
High
Next
1. Use
to call up the edit mode.
2. Adjust the desired state in case of faults with
3. Confirm the selection with
Easy Setup
High Alarm Value
21.000 mA
Next
4. Use
to call up the edit mode.
5. Use
/
/
to set the alarm current.
6. Confirm the selection with
Zero point adjustment of the flowmeter
NOTICE
Prior to starting the zero point adjustment, make sure that:
— There is no flow through the sensor (close valves, shut-
off devices etc.).
— The sensor is completely filled with the medium to be
measured
Easy Setup
Auto Zero
Next
— Use
to start automatic adjustment of the zero point for
the system.
Change from two to one column
44 CI/FSV/FSS/430/450-EN Rev. G | VortexMaster FSV430, FSV450 SwirlMaster FSS430, FSS450
Edit
.
Easy Setup
Low Alarm Value
Edit
Next
.
Edit
Configuration of the low flow cut-off
Easy Setup
Low Flow Cutoff
Next
1. Use
to call up the edit mode.
2. Use
/
/
.
cut-off.
3. Confirm the selection with
Menu
Easy Setup
3.600 mA
Edit
Exit
Once all parameter have been set, the main menu appears
again. The most important parameters are now set.
4. Use
to switch to the process display.
0.000 %
Edit
/
to set the desired value for the low flow
.
Select