Quantel Ultra Big Sky Laser Series User Manual page 89

Table of Contents

Advertisement

Interlock
ICE Front Panel
Value
Message
IQS 10000000 Q-S disabled
IQS 01000000 Q-S disabled
coolant temperature
under limit
IQS 00001000
Q-S disabled iQS_4
IQS 00000100
Q-S disabled iQS_3
IQS 00000010
Q-S disabled iQS_2
IQS 00000001
Q-S disabled iQS_1
|
Ultra
Possible Cause
This interlock does
Contact Quantel Customer Service for
not prevent the laser
repair.
from firing. It prevents
Q-Switch activation
before the safety
timeout expires.
This interlock does
Allow the coolant to warm up.
not prevent the laser
Contact Quantel Customer Service to
from firing. It prevents
inquire about increasing the minimum
Q-Switch activation
Q-Switch operating temperature.
when the coolant is
below the minimum
operating temperature.
Contact Quantel Customer Service.
Solution/Suggestion
Troubleshooting |
Contents
Description
This interlock is not used. If this value
appears, contact Quantel Customer
Service for repair.
To protect optics and ensure reliable laser
performance, the coolant temperature must
meet a required minimum or Q-Switch
operation is locked out. This interlock
monitors the coolant temperature using a
sensor located in the coolant reservoir.
These interlocks are not used. If they occur,
they indicate a problem.
Version M 22-April-2012