Quantel Ultra Big Sky Laser Series User Manual page 47

Table of Contents

Advertisement

Feature
Operation
Interlock Fault Byte 2, query
Note: For more information about the
interlocks, refer to
on page
78.
Example: IF2
IF2 00 00 01 00
indicates low coolant flow
Status
Commands
Interlock Fault Byte 3, query
(continued)
Example: IF3
indicates simmer stopped
|
Ultra
Command
IF2
Troubleshooting
IF3
IF3 00 10 00 00
Response
Description
format
IF2 ab cd ef gh
The response string contains ASCII-encoded 1's or 0's
indicating that either a fault or okay condition exists.
a:
"1" if heater thermostat is open
b:
"1" if charger temperature is over the
maximum temperature setting
c:
"1" if coolant temperature is under the
minimum temperature setting
d:
"1" if coolant temperature is over the
maximum temperature setting
e:
"1" if coolant level is low
f:
"1" if coolant flow is low
g:
"1" if charger, coolant, or harmonic
generator temperature is below the
minimum setting
h:
"1" if flashlamp power setting is too high
IF3 ab cd ef gh
The response string contains ASCII-encoded 1's or 0's
indicating that either a fault or okay condition exists.
a:
"1" if no end of charge prior to fire order
(PSU charge error)
b:
"1" if voltage is over setting
c:
"1" if no simmer sensed
d:
"1" if external flash signal frequency is
too low
e:
"1" if external flash signal frequency is
too high
f:
"1" if there is a capacitor discharge
problem
g:
"1" if there is simmer timeout
h:
"1" if the PIV slave indicates master unit
has interlock and vice versa
| 43
Software |
Contents
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
Version M 22-April-2012