38 |
Contents
| Software
Feature
Operation
Fire Q-Switch, set
Example: CC
Fire Q-Switch Single Shot
Example: OP
Stop Q-Switch
Q-Switch
Example: CS
Commands
Q-Switch Delay, query
Example: W
Q-Switch Delay, set
Example: W144
Version M 22-April-2012
Command
CC
fire auto qs
OP
fire auto qs
CS
fire auto
W
delay 123 μs
Wnnn
where nnn is
delay 144 μs
the Q-Switch
delay in
microseconds
Response
Description
format
state string
Starts the Q-Switching. The response to this command is the
state string.
state string
Initiates a single Q-Switch. The response to this command is
the state string.
Note: This functions only when Q-Switch mode is set to INT.
state string
Disables the Q-Switch. The response to this command is the
state string.
delay nnn μs
Queries or sets the Q-Switch delay setting. This is the
amount of time, in microseconds, between the flashlamp
where nnn is the
Q-Switch delay in
activation and the Q-Switch pulse when the Q-Switch is in
microseconds
Internal sync mode. The max-delay command ("WMX") sets
the maximum allowable delay, and the min-delay command
("WMN") sets the minimum allowable delay.
|