44 |
Contents
| Software
Feature
Operation
Q-Switch Interlock Byte, query
Example: IQS
indicates coolant is too cold
Status
Commands
(continued)
First Interlock
Example: IF
Cooling Group Temperature
Example: CGT temp.
indicating the coolant temperature is 45°C
Harmonic Generator Temperature
Example: HGT temp.
Temperature
d
Commands
indicates the harmonic generator temperature is
75°C
Charger/Simmer Temperature
Example: CST
indicates the charger/simmer temperature is 55°C
Version M 22-April-2012
Command
IQ
IQS 01 00 00 00
IF
IF2 00 00 01 00
CGT
CG 45 d
HGT
SHG 75
CST
temp.
CS 55 d
Response
Description
format
IQS ab cd ef gh
The response string contains ASCII-encoded 1's or 0's
indicating that either a fault or okay condition exists.
a:
"1" if in the 8-second forced delay after
flashlamp starts
b:
"1" if coolant temperature is too low
c:
"1" if Q-Switch timeout
d:
"1" if shutter is closed
e:
not used
f:
not used
g:
not used
h:
not used
??? ab cd ef gh
Queries the interlock that caused a laser shutdown. The
response is that of IF1, IF2, IF3 or IQ depending on which
one caused the shutdown. In the case of multiple causes,
priority is from IF1 down to IQ.
temp. CG nn d
Queries the temperature of the coolant in degrees Celsius
temp. SHG nn d
Queries the temperature of the harmonic generator in
degrees Celsius
temp. CS nn d
Queries the temperature of the charger/simmer controller in
degrees Celsius
|
Ultra
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not