System Info Menu
From the Main menu, scroll to the "System Info" item and press the "Enter
Menu" button to access the "System Info" menu items:
Main Menu
>Temp
34.1°C
serial link
OFF
Time-out
OFF
ICE450/BRIO
version 3.05
option HG
OFF
Interlocks
Pump
ON
Main Menu:
To return to the Main menu scroll the cursor to the
"Main Menu" line and press the "Enter Menu" button.
Temp:
This item displays the coolant temperature measured within
the reservoir. The "Temp" item also provides a submenu when
the "Enter Menu" button is pressed. The "Temp" menu shows the
following items:
System menu
>CG temp
HG temp
CS temp
Tmin fl
Tmax fl
Tmin qs
heater
flow
level
|
Ultra
34.1°C
37.1°C
43.2°C
30°C
55°C
35°C
44%
1.805 lpm
OK
Temp Submenu items:
System menu: Use the "Enter Menu" button to select this item to
return to the System Info menu.
CG temp: Displays the Cooling Group (coolant) temperature
within the reservoir.
HG temp: Displays the temperature of the Harmonic Generator, if
installed.
CS temp: Shows the temperature of the Charger-Simmer
electronics and is useful for diagnostics.
Tmin fl: Shows the minimum allowable coolant temperature for the
flashlamp to operate. This is set at the factory.
Tmax fl: Shows the maximum coolant temperature. This is a
factory setting. If the measured temperature is above this, an
interlock condition occurs and the flashlamp will not operate.
Tmin qs: Shows the minimum coolant temperature for Q-Switch
operation. This is set at the factory.
heater: Shows the duty cycle of the heater element that warms
the coolant.
flow: Shows the flow rate of the coolant.
level: Shows "LOW" if the coolant level in the reservoir is too low.
serial link:
Set to enable (ON)/disable (OFF) the RS-232 serial
communications link. The default setting is ON at power-up, enabling the
RS-232 communications link. For safety reasons, when any button on the
Remote Box is pressed, the serial link is disabled (OFF) giving priority to
the Remote Box.
To re-enable the serial communications, use the Remote Box to select
"serial link" and set it to ON.
Note: If you are unable to communicate with the ICE450 using the serial
link, check the serial link setting from the System Info menu.
| 25
Operation |