Lamp Insertion
1. Clean the new lamp with methanol or isopropyl alcohol.
Note: While the lamp is very slightly damp from the methanol or
alcohol, it slides more easily through the O-rings seals inside the Laser
Head.
2. Manually slide the lamp into the laser. Press with straight and
even pressure to avoid breaking the flashlamp. Do not tilt the
Laser Head during installation of the flashlamp. You will feel
the lamp make a solid contact when completely installed. If
you are unsure ifthe lamp is fully seated, pull the lamp back out
approximately 1/2" [12mm] and reseat it.
CAUTION: Operation of the pump with a flashlamp that is not
fully seated will flood the Laser Head and cause permanent
laser damage. Operation of the pump with a flashlamp that
is not properly installed will void the warranty. If there is any
doubt about this lamp insertion process, please call Quantel
USA Customer Service Department.
3. Replace the lamp access cover by pushing and rotating it
clockwise. Verify that the O-ring seal is properly installed between
the lamp access cover and the Laser Head.
4. Purge the Laser Head with dry nitrogen as described in
Nitrogen Purge on page
|
Ultra
76.
Resetting the Shot Counter
After replacing the flashlamp, follow these steps to reset the shot counter
using the Flashlamp menu from the Remote Box:
1. From the Main menu, select Flashlamp and press the "Enter Menu"
button to show the Flashlamp menu.
2. Scroll the cursor to the "cu" item.
3. Press the "Enter Menu" button.
This will set the value to zero.
| 75
Maintenance |
Contents
Clean the new lamp with
methanol or isopropyl
alcohol before inserting it
Enter Menu
button
Version M 22-April-2012