| Troubleshooting
Interlock
ICE Front Panel
Value
Message
IF2 00001000 low coolant level
ICE cooling
IF2 00000100 low coolant flow
IF2 00000010 defective temp.
sensor
IF2 00000001 power too high
NOTE: To clear an interlock message, press the "Cursor
Up" or "Cursor Down" button on the Remote Box.
Version M 22-April-2012
Possible Cause
• System is low on
• Ensure coolant is properly filled after
coolant
draining the coolant system. Follow the
instructions in
• Changing to longer
• If operating the ICE450 horizontally,
coolant lines requires
ensure the coolant fill port is below the vent
additional coolant
port. See front and back panel labels.
• The ICE450 is
There may be a leak or the level-detect
horizontal with the
switch may be faulty. Call Quantel
wrong side facing up
Customer Support.
• Coolant line is
• Examine the coolant lines. Ensure they
obstructed preventing
are not kinked or otherwise obstructed.
proper flow
• Ensure the coolant line connectors are
• Flow sensor is
fully engaged and locked.
damaged or stuck.
• Ensure the pump is toggled ON from the
Remote Box.
The pump or the flow sensor may require
repair. Notify Quantel Customer Service.
One or more of the
• Ensure all temperatures are above 18 °C.
three feedback
• If the system does not include a harmonic
thermistor circuits
generator, make sure the HG option is
(coolant, charger/
turned off from the System Info menu.
simmer, and harmonic
Note: A reading of 17.9°C indicates a
generator) is faulty or is
temperature feedback open circuit. Notify
running too cold.
Quantel Customer Service in this case.
Voltage or frequency
Reduce voltage, energy, or frequency until
settings are too high,
the power is less than 450W. Verify this
resulting in power too
and save the configuration.
high.
Solution/Suggestion
Maintenance on page
68.
|
Ultra
Description
A coolant level float style switch is located
in the coolant reservoir. When this interlock
is sensed, in addition to the displayed
message on the Remote Box, the reservoir
backlights will flash.
A flow-rate sensor located in the ICE450
coolant loop triggers an interlock condition
when the flow rate is less than 2.2 lpm
(+/- 10%).
A defective temperature sensor interlock
appears if any one of the three temperature
feedback signals (CG, CS or HG) is
abnormal.
The ICE450 calculates the power delivered
to the lamp. If the power per shot is more
than 450W, this interlock is activated.
Note: Under normal conditions you are
prevented from adjusting the settings too
high. The interlock prevents serious problems.