Operation
Flashlamp Energy, query
Example: EJ
Flashlamp Energy, set
Example: EJ600
Flashlamp Frequency, query
Example: D
Flashlamp Frequency, set
Example: D2200
Flashlamp Count, query
Flashlamp
Commands
ct LP0000012345
Example: F
(continued)
User Flashlamp Count, query
Example: UF cu LP0000012345
Reset User Flashlamp Count
Example: UF0 cu LP0000000000
Flashlamp External Trigger Delay Bypass, query
Example: BYPASS
Flashlamp External Trigger Delay Bypass, set
Example: BYPASS1
|
Ultra
Command
EJ
energy 06.77 J
EJnnnnn
where nnnnn is
energy 06.00 J
in hundredths
of Joules.
D
freq 021.00Hz
Dnnnnn
where nnnnn
freq 022.00Hz
is the desired
frequency in
hundredths of
hertz
F
.
UF
UF0
BYPASS
bypass:off
BYPASSn
where n is
bypass:on
0 for off,
1 for on
Response
Description
format
energy xx.yy J
If the ICE450 has been configured to allow energy settings
(using the "FSM" factory command), this command queries
where "xx.yy" is the
or sets the flashlamp energy. The energy is the electrical
energy in joules
energy stored in the flashlamp capacitor, calculated as
½CV².
freq xxx.yyHz
Queries or sets the flashlamp firing frequency. This is
the rate at which the flashlamp fires when in Internal
where "xxx.yy" is the
Synchronization mode. When the flashlamp is in External
present frequency
setting in hertz.
Sync, the external trigger frequency compares with this
setting.
ct LPnnnnnnnnn
Queries the total number of flashlamp firings since shipment
from the factory.
cu LPnnnnnnnnnn
Queries the number of flashlamp firings since the resetting
of the count. Reset the flashlamp count with every flashlamp
replacement.
cu LP0000000000
Resets the number of flashlamp firings. Send this command
after replacing the flashlamp.
bypass: nnnn
Queries or sets the external flashlamp trigger delay bypass.
When BYPASS is set to 0 (off), there is a 500 μsec delay
where "nnnn" is "off"
between the rising edge of the external tirgger input and the
μ
(500
sec delay) or
rising edge of the flashlamp synch internal trigger. When
μ
"on" (0.5
sec delay).
BYPASS is set to 1 (on), there is only a 0.5 μsec delay.
The response to a query is the current state of the bypass
(on or off). The BYPASS command only takes effect when
transitioning from Stop to Fire.
Note: The delay bypass is only valid if Q-Switch Trigger
mode is set to external (QE). If the Q-Switch trigger mode is
set to internal (QI) there will be a 500 μsec delay.