Interlock
ICE Front Panel
Value
Message
IF3 10000000 PSU charge error
IF3 01000000 voltage over setting
IF3 00100000 simmer stop
NOTE: To clear an interlock message, press the "Cursor
Up" or "Cursor Down" button on the Remote Box.
|
Ultra
Possible Cause
The charger did not
• Reduce the frequency or voltage/energy
have time to fully
of the firing signal.
charge before another
• Verify the Laser Head I/O cable connectors
Fire signal was
are fully inserted and tightened.
received.
• Inspect the high voltage/high current
contacts for damage caused by arcing.
Due to firmware priorities, if the Charger/
Simmer overtemperature interlock trips
during flashlamp operation, a PSU charge
error may display on the Remote Box if
the charge cycle is disabled before the
microprocessor is notified. Refer to
IF2 01000000 for details.
If the error persists, please contact Quantel
Customer Service.
This indicates a
problem with the
charging system.
The simmer current
• Verify that the connections to the Laser
either failed to be
Head are secure.
established, or it failed
• View the flashlamp shot count. Replace
to stay active.
the flashlamp if necessary.
Solution/Suggestion
CAUTION: This is a potentially
serious safety issue. Please contact
Quantel Customer Service.
Troubleshooting |
Contents
Description
The system controller must receive an
"end of charge" feedback signal from the
charger, prior to the Fire order for flashlamp
discharge. If not, a PSU charge error
interlock occurs.
The "end of charge" feedback indicates the
PFN capacitor has charged to the desired
voltage.
This interlock indicates that the Pulse
Forming Network (PFN) capacitor voltage
actual value exceeds the programmed
voltage by 5% or more.
The flashlamp simmer current is constantly
monitored. If the simmer current fails to
establish or "blows out", then this interlock
is activated.