Quantel Ultra Big Sky Laser Series User Manual page 46

Table of Contents

Advertisement

42 |
Contents
| Software
Feature
Operation
Status Word, query
Example: WOR I 0 F 2 S 1 Q 6
Indicates there are no interlocks (I 0), the
flashlamp is in Internal sync mode and is
running (F 2), the simmer is established
(S 1), and the Q-Switch is in External
sync and is running (Q 6)
Status
Commands
(continued)
Interlock Fault Byte 1, query
Example: IF1
indicates Laser Head housing is opened
Note: For more information about the
interlocks, refer to
on page
78.
Version M 22-April-2012
Command
WOR
IF1
IF1 00 01 00 00
Troubleshooting
Response
Description
format
I a F b S c Q d
This command is included for compatibility with the Brilliant
ICE450. It queries for a code that indicates the state of the
laser. The response is:
I a F b S c Q d, where:
a: "0" = No interlocks present; "1" = Interlocks present
b: "0" = The flashlamp is in Internal sync mode and is not
running; "2" = The flashlamp is in Internal sync mode and is
running; "4" = The flashlamp is in External sync mode and
is not running; "6" = The flashlamp is in External sync mode
and is running.
c: "1" = Simmer is established; "0" = Simmer is not
established
d: "0" = The Q-Switch is in Internal sync and is off; "1" =
The Q-Switch is in Internal sync and is in single mode; "2"
= The Q-Switch is in Internal sync and is running; "4" = The
Q-Switch is in External sync and is off; "6" = The Q-Switch is
in External sync and is running.
IF1 ab cd ef gh
Queries Interlock Fault Byte 1. The response string contains
ASCII-encoded 1's or 0's indicating either a fault (1) or okay
(0) condition exists. The format of the string is as follows:
IF1 ab cd ef gh, where the letters correspond to the
following interlocks:
a:
"1" if e-stop button is pressed
b:
"1" if bnc interlock is open
c:
"1" if Laser Head thermostat is open
d:
"1" if Laser Head housing switch is open
e:
"1" if ICE450 housing switch is open
f:
"1" if there is an Internal bus error
g:
"1" if there is an External bus error
h:
"1" if flashlamp timeout
|
Ultra
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not
0 if not

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ultra ice 450

Table of Contents