Visual Indicators
The intended operator's position for correctly perceiving a visual signal and its priority is 1 meter
(3.3 feet).
The following table describes conditions and visual indicators.
Condition
Pulse Waveform Signal is
inadequate
Sensor Fault (i.e., sensor
disconnect, failure, or
incompatibility with the monitor)
Inadequate Signal (sensor
misalignment, ambient light, nail
polish/artificial nails, etc.)
Inadequate SpO
or pulse rate
2
data (excessive motion or erratic
heart rate, etc.) more than
20 seconds
Pulse rate data not updated for
more than 30 seconds
Low Battery
Critically Low Battery
14
Visual Indication
• Pulse Quality LED blinks red
• Pulse Quality LED blank
• Dash (-) appears in the leftmost position of the SpO
display
• SpO
and Pulse Rate numeric displays freeze for
2
10 seconds
• After 10 seconds a dash (-) replaces the SpO
Pulse Rate numeric displays.
• Pulse Quality LED blinks
• Dash (-) appears in the leftmost position of the SpO
display
• SpO
and Pulse Rate numeric displays freeze for
2
10 seconds
• After 10 seconds a dash (-) replaces the SpO
Pulse Rate numeric displays.
• Dash (-) appears in SpO
• Pulse rate numeric display becomes dashes
• SpO
and Pulse Rate displays blink.
2
• SpO
and pulse rate displays are blinking dashes
2
• Pulse Quality LED is solid red or amber
2
2
and pulse rate displays
2
Visual Indicators
2
and
2
and