Equipment Response Time

If the signal from the sensor is inadequate, the last measured SpO
10 seconds and are then replaced with dashes.
Standard/Fast Averaged SpO
Standard/Fast Averaged Pulse Rate
Example - SpO
Exponential Averaging
2
SpO
decreases 0.75% per second (7.5% over 10 seconds)
2
Pulse Rate = 75 BPM
Specific to this example:
• The response of the 4-beat average is 1.5 seconds.
24
SpO
Values
2
Pulse Rate Values
Equipment Delays
Display Update Delay
SaO2 Reference
100
95
90
85
80
75
70
65
60
and pulse rate values freeze for
2
Average
4 beat exponential
2
Response
4 beat exponential
Delay
1.5 seconds
4 Beat Average
Time in seconds
Technical Information
Latency
2 beats
Latency
2 beats