DIReCTIOnS fOR USe

PRePARATIOn fOR USe

CAUTIOn
Unpack and inspect all components. Identify the components and their use with the pictures
located in the Product Description Section of this manual.
All components are shipped in a clean but not sterile condition. If the Spine Positioning
System II will be used for an indicated surgical procedure, be sure to disinfect the com-
ponents prior to use. Disinfectants that can be used are described in the Cleaning &
Disinfecting Section of this manual.

fACe ReST PlATfORM ADjUSTMenT

Step 1 - Open cam
Step 4 - Continue to close cam.
Do not overhang the platform frame beyond the Warning line
on the frame.
Step 2 - Grasp platform and raise to
desired position
Step 5 - Close cam to the final position.
Step 3 - Begin to close the cam making
sure that the small locking pins enter
corresponding positioning holes (Minor
platform "rocking" may be necessary for
the pins to enter the holes). Do not force
the cam to close
Step 6 - Double check platform by ap-
plying downward force to ensure your
face rest platform is securely locked.