Symbols & Units Of Display; Resistance Measurement - Lutron Electronics MO-2014 Operation Manual

10 amperes max. test current micro-ohm meter
Table of Contents

Advertisement

6 -2 Symbols & units of display
Symbol &
Unit
uΩ ,mΩ ,Ω
.)))
HOLD
GOOD
FAIL
COMP
L
H I
LO

6 -3 Resistance Measurement

1) Power on the instrument by rotate the "Power /
Range rotary switch" ( 3-10, Fig. 1 ) to exit the " OFF "
position and then select measuring range from 60000 uΩ
to 6000 Ω according to your requirement.
Note :
* Always select the highest range ( 6000 Ω
) if you don't know the resistance value of
the resistor and then select lower range
in sequence.
* At 60000 uΩ ( 10A )must use DC 9V
adapter from socket ( 3-17, Fig. 1 ).
* Over range LCD shows " ----- ".
* Can't measure the resistance value, LCD
will show 0.0000, 0.000 or 0.00.
Function
Ohm unit
Appears on the "BEEP" function have
started.
Appears on the"HOLD" function have
started.
It is pass Q.C. Appears on finished
compare operate.
It is higher or lower than maximum or
minimum acceptance value . Appears
on finished compare operate.
Appears on the"COMPARE" function.
Appears on the"INDUCTIVE" mode.
Appears to showing H I limit.
Appears to showing LO limit.
6