OAKWORKS 300 Series User Manual page 5

Procedure chair
Hide thumbs Also See for 300 Series:
Table of Contents

Advertisement

IMPORTANT SAFETY INSTRUCTIONS
The arm rests are not designed to be used as patient restraints. Use appropriate patient restraints.
The chair should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the chair should be
observed to verify normal operation in the configuration in which it will be used.
The chair is designed to be a stand alone chair. This chair must not be modified or incorporated into any other equipment.
As with any moving mechanism there are potential pinch points around and underneath the chair. It is the responsibility of the
operator of this equipment to insure that bystanders are not in the area below or around this equipment during operation.
Proper operation of this equipment is very important for the safety of the operator, patient, and any other individuals in the area of
this equipment. Directions for use of this equipment are described in this manual. The operator should read these sections carefully.
Weight Limit: (patient and accessories) 550 lbs./250 kg. Do not exceed.
Be certain that the chair is completely lowered without any tilt being present prior to discharging an ambulatory patient. The
patient may lose balance and fall.
This chair is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
When lowering the chair or using the Trendelenburg functions, make sure there is nothing underneath the chair that can impede
motion (like stools, cabinets, accessory parts, cleaners, etc.).
Use this furnishing only for its intended use as described in these instructions. Do not use attachments not recommended by the
manufacturer.
Close supervision is necessary when this furnishing is used near children or disabled persons.
The Oakworks Procedure Chair has two lifting columns that raise and lower the chair and three actuators that allow the client
to be positioned for both client and therapist comfort. Each of these devices has a duty cycle of approximately 10%, meaning 2
minutes of operation within a 20 minute period. Exceeding this duty cycle can damage the equipment.
Do not sit on the Backrest or Leg Rest.
WARNING
To reduce the risk of burns, fire, electric shock or injury to persons:
1. Unplug this furnishing from the electrical outlet before cleaning.
2. Unplug from outlet before adding or removing parts.
3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or
damaged, or dropped into water. Contact Oakworks Customer Service before use.
4. Keep the cord away from heated surfaces.
5. Never drop or insert any object into any opening.
6. Do not use outdoors.
7. Do not operate where aerosol (spray) products are being used or where oxygen is being administered.
8. The maximum amperage for the Power Outlet is 3 amps. Connecting equipment that exceeds this rating will damage the
system. Only connect low current devices. Do not connect an extension cord or power strip to the Power Outlet. Do not
touch equipment connected to the Power Outlet and the patient at the same time. Any electrical accessories used with this table
must comply with approved medical standards. Equipment relying only on Basic Insulation (only one means of electrical protection
for the patient / practitioner) must not be used.
DANGER
Risk of electric shock - Connect this furnishing to a properly grounded outlet only. See Grounding Instructions in this manual.
Electrical Shock Hazard. The power supply/control module is located under the chair seat. No user serviceable parts are inside the
control box. Refer servicing to qualified personnel. Unplug wall plug prior to contact with any cables connected to the power
supply.
2

Advertisement

Table of Contents
loading

Table of Contents