I.V. Pole (058-737)
The optional I.V. pole must be installed on the patient's right side of the table.
Installing the I.V. pole:
1. Using a 7/16" wrench, remove the two 1/4-20" hex head screws from the end of the tabletop
frame on the side of the table to which the I.V. pole will be installed.
2. Position the I.V. pole support bracket against the side frame so the two pre-drilled screw holes
in the bracket align with the screw holes on the tabletop frame. The I.V. pole should face up
and toward the head end of the table.
3. Using a 7/16" wrench, install the two screws removed in step 1 through the I.V. pole mounting
bracket. Tighten the screws to secure.
4. Insert the I.V. pole into the I.V. pole mounting bracket. Ensure that the I.V. pole passes through
both the top and bottom holes in the mounting bracket to ensure I.V. pole stability.
Figure 12. Installing the I.V. pole.