UROLOGY & BRACHYTHERAPY C-ARM TABLES™
INSTRUCTIONS FOR USE
058-800
058-805
058-810
058-815
FN: 20-324-CLR 11/20

Advertisement

Table of Contents
loading

Summary of Contents for biodex 058-800

  • Page 1 UROLOGY & BRACHYTHERAPY C-ARM TABLES™ INSTRUCTIONS FOR USE 058-800 058-805 058-810 058-815 FN: 20-324-CLR 11/20...
  • Page 2 Tables 058-800, 058-805, 058-810, and 058-815. Additional information and resources are available upon request or directly from the Biodex website, www.biodex.com/tables800. If the desired information is not found, contact a local distributor or Biodex directly at supportservices@biodex.com Thank you, Biodex Medical Systems, Inc.
  • Page 3: Table Of Contents

    Table of Contents   Definition of Symbols ........................ 4   Product Certifications and Classifications .................. 5   Before Proceeding ........................6   Important Safety Information ....................7     Introduction ........................8   Intended Use ......................... 8   Indications for Use ......................8  ...
  • Page 4: Definition Of Symbols

    Definition of Symbols The following symbols and their associated definitions are used and implied throughout this manual.
  • Page 5: Product Certifications And Classifications

    Product Certifications and Classifications The Urology & Brachytherapy C-Arm Tables have received the following certifications, and fall within the following classifications: IEC60601-1:2005 (Third Edition) + CORR1:2006+Corr 2:2007 + A1:2012 (or IEC 60601- ● 1:2012 reprint) ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010 ● CAN/CSA C22.2 No: 60601-1:14 ●...
  • Page 6: Before Proceeding

    Before Proceeding NOTE: The warnings, cautions and instructions provided in this manual must be read, followed and kept available for consultation at all times. Observing the information, instructions and procedures presented throughout this manual is essential for using this product both properly and safely. SPECIFIC CAUTIONS Allow only qualified, trained personnel to operate or service this product.
  • Page 7: Important Safety Information

    Before Proceeding section of this manual. This product should be used only as specified in the operation manual. CAUTION: Biodex devices are designed for use in a client environment. For product specifications, refer to the Table of Contents. This medical electrical equipment requires special precautions regarding EMC and must be assembled and placed into service according to EMC information provided in this manual.
  • Page 8: Introduction

    1. Introduction Intended Use The Urology and Brachytherapy C-Arm Tables feature motorized actuation of height, X-Y tabletop, lateral roll and Trendelenburg motions, and is supported by a full line of optional accessories. Both AC and battery power are standard. When operating on battery, there are no power cords to catch on feet or block table wheels.
  • Page 9 Figure 1.1. Parts and adjustment mechanisms. Standard Parts and Adjustments: 1. Hand-Held Controller 2. Foot Controller 3. Fowler Back 4. Drain Bag Holder and Disposable Bag 5. Left and Right Knee Crutches 6. X-Ray Grid and Film Cassette Tray 7. Rail-Mounted I.V. Pole 8.
  • Page 10: Operating Instructions

    2. Operating Instructions CAUTION: Refer to instruction manual before operating or servicing system. NOTE: This table can be operated with the power cord connected to an outlet or by using the battery. The battery recharges automatically while the table is plugged into an outlet even if the table is being used.
  • Page 11: Fowler Positioning

    Figure 2.1. Locking Brakes and Swivel Casters Fowler Positioning The Fowler Back, manually adjustable to an 80-degree angle (20°, 30°, 40°, 50°, 60°, 70°, 80°) incorporates a touch-handle lock/unlock mechanism for effortless patient positioning. CAUTION: The Fowler Back must be supported as it is being raised or lowered. The back must lock into one of the eight provided positions.
  • Page 12: Using The Foot Controller

    To adjust the table position, press and hold the appropriate switch on the hand-held controller. The tabletop will move or rotate in the direction indicated by the switch. Release the switch when the tabletop achieves the desired position. Figure 2.2. Hand-Held Controller Using the Foot Controller The Foot Controller activates all motions.
  • Page 13: Disposable Skirt (Optional)

    Slide the Extension Out of the Table. Disposable Skirt (Optional) The Biodex Urology & Brachytherapy C-Arm Tables (models 058-800 and 058-810) can be ordered with the optional Plastic Sanitary Skirts to protect the table bellows from patient blood and other liquids. The table has plastic clips located immediately above the bellows.
  • Page 14: Pole (Optional)

    3. Remove and replace the plastic skirt for each new patient. NOTE: Additional Bellows Skirts (part #056-802) can be ordered from Biodex Medical Systems, Inc. Figure 2.5. Plastic Skirt Clip Open Position. Figure 2.6. Slide the plastic skirt underneath and lower the clip to hold the skirt in place.
  • Page 15: Arm Boards (Optional)

    To adjust positioning of the crutch pads, loosen the crutch pad locking levers and tilt the pads as required. Figure 2.7. Knee Crutches Properly Installed. Arm Boards (Optional) Standard or Dura Board Arm Boards are optional. Install the rail clamps onto the accessory rails and then insert the arm board posts into the clamps and tighten to secure.
  • Page 16: Drain Pan (Optional)

    Figure 2.8. The Drain Bag Holder and Flexible, Disposable Drain Bag Properly Installed. Drain Pan (Optional) The stainless steel drain pan and wire mesh screen are optional. To insert the drain pan, open the vertical doors at the foot end of the table (the doors pivot toward the outside of the table. Pull out on the drain pan support bars until they are fully extended.
  • Page 17: Radiographic Film Cassette Tray

    “drops in” past the stop notch mechanism. The x-ray grid should remain in the table at all times except when the table is being used for fluoroscopic purposes. Radiographic Film Cassette Tray The optional Radiographic Film Cassette Tray can be inserted or removed from either side of the table.
  • Page 18: Maintenance And Cleaning

    3. This table is rated to support patients weighing up to a maximum of 500 pounds. 3. Maintenance and Cleaning The Biodex Urology & Brachytherapy C-Arm Tables are virtually maintenance free. By following the instructions below at suggested time intervals, or as often as necessary, the table(s) will remain in “like new” condition.
  • Page 19: Specifications

    The disposal of equipment must be in accordance with the respective national regulations. Wear parts are considered hazardous waste! After being replaced, wear parts must be disposed of according to country-specific waste laws. 4. Specifications Dimensions: 84.5" l x 25.5" w (214.6 x 64.8 cm) with OR accessory rails Tabletop: 53"...
  • Page 20 Optional 056-463 Grid, X-ray 056-862 Battery and Charging Unit, 115 VAC 056-863 Battery and Charging Unit, 230 VAC 056-850 IV Pole, Rail Mounted 056-851 Drain Bag Frame 056-853 Crutch, Knee, Simplicity (pair) 058-860 Drain Pan, Stainless Steel 058-862 Stirrups, PAL with Boot Pad (pair) 058-863 Arm Board, Dura Board with Pad (pair) 058-865...