Pentair HAFFMANS RPU-352 Instruction Manual page 84

Redpost pu monitor
Table of Contents

Advertisement

2
T6
This recording run can't be saved as a template. In order to save a recording run as a template
the following conditions must all be met:
Channel 1 must have been active in the P.U. monitor.
The run must contain no off-scale readings on any active channel.
The run must contain at least 25 records.
This means that a run must be more than 4.2 minutes long with a recording interval 10
seconds or more than 12 minutes long with a recording interval of 30 seconds and pro-
rata for other intervals.
There must have been no communications errors during the data transfer from the P.U.
monitor to the RPC-80.
All the expected data must have been received from the P.U. monitor.
T7
Saving this template failed when the memory became full.
Save the template in the place of a template that is no longer required.
There may still be room in the memory for a smaller template.
Two sizes of template storage memory are available for the RPC-80, one with 127 sectors and
the other with 511. If you have the smaller type and full template memory becomes a problem
then you may wish to have your RPC-80 upgraded.
T8
Saving this template failed when a communications error occurred
Connects the lead between the P.U. monitor and the RPC-80.
Never unplug this lead until the "Template Saved OK" message shows.
When a communications error occurs the partly saved template is deleted and another
an other chance will be offered to save the run as a template.
ATTENTION!
Page 84 of 112
It is possible that an error could be caused by very severe interference
from other nearby equipment or on the mains wiring.
HAFFMANS RPU 352, RPU 353, RPC 80 RPC 50
RPU 352, RPU 353, RPC 80, RPC 50