P-Messages - Pentair HAFFMANS RPU-352 Instruction Manual

Redpost pu monitor
Table of Contents

Advertisement

2
ATTENTION!
M3
The memory that stores the recorded templates for the RPC-80 has become corrupted and all
the data in it has been lost. The memory has been tested and found to be operating correctly.
The saved templates will have all been deleted and templates turned off.
The most likely reason for the memory loss is that the back-up battery power was lost for a short
time. This may have been due to a fault in the battery or the connector. Check that the battery
connector has not been disconnected by accident.
M4
The memory that stores the recorded templates for the RPC-80 has become corrupted and the
data in it has been lost. The memory has been tested and found to be faulty, and has been
taken out of service.
The templates will have been turned off and templates cannot be used.
If this fault occurs it will be necessary to replace the template memory. Return the instrument for
repair. The RPC-80 will continue to function normally without the template facility.

26.4 P-MESSAGES

All messages starting with a "P" are messages concerning the printing.
P1
The printer is not responding to the instructions. See the trouble-shooting section of your printer
manual.
Check that the printer is "on line"
Do not try to start printing before the "on line" lamp is lit permanently.
Check for sufficient paper
Check for a paper jam.
Check the data lead connecting the RPC-80 and the printer.
100 Series monitors
If you start to print as soon as the "PRINT" button appears, the data from the monitor is still
being received and the on/off lamp on the monitor is on steadily. If the paper runs out or any
other printer fault causes a hold up of more than about 6 minutes then the monitor will return to
idle with the on/off lamp flashing every 2 seconds. The print run may then be incomplete. Once
the printer fault is cleared you should abandon the run by disconnecting the monitor from the
RPC-80 then reconnect it and restart the print run.
Page 82 of 112
Enter the correct time and date that the back-up battery was fitted
followed by the serial number of the instrument. Until these items have
been correctly entered the fault will recur on each power up.
HAFFMANS RPU 352, RPU 353, RPC 80 RPC 50
RPU 352, RPU 353, RPC 80, RPC 50