Battery life
From a fully charged battery
Example (a) 2 maximum runs
Example (b) 15 typical runs
Assumes final recording is completed within one week of the first
Dimensions
(Length x Width x Height)
In frame
Weight
9.25 kg
Dimensions and weight are inclusive of carrying frame, can holder and can pressure fitting
Materials
Stainless steel, toughened glass and PET-P plastic.
Calculation of pasteurisation units (PU) can take place in the RPU-352, PRU-353
Accuracy
Range
Resolution
* Note that a temperature measurement error of ± 0.25 ºC represents ± 8% error in PU value.
Page 10 of 112
HAFFMANS RPU 352, RPU 353, RPC 80 RPC 50
Record 4 hours, display (with
backlight) 20 minutes
Record 40 minutes, display (with
backlight) 5 minutes
380 x 175 x 230 mm
Including
frame
Better than ± 0.001% for calculation
alone *
0 to 9999.9
PU
0.1 PU
RPU 352, RPU 353, RPC 80, RPC 50