Recording Interval; Equipment Number; Mains Frequency; Leaving Configuration - Pentair HAFFMANS RPU-352 Instruction Manual

Redpost pu monitor
Table of Contents

Advertisement

2

13.11 RECORDING INTERVAL

This screen allows you to select the time interval between recordings. You can
choose any value between 2 and 60 seconds. The factory setting is 10 seconds.

13.12 EQUIPMENT NUMBER

This screen allows you to set your own equipment number or identification for the
instrument. This can consist of up to 15 letters or numbers and will be printed
with the results by the Redpost RPC-80 printer/charger. This identification may
be required for ISO9000 schemes. Only upper case Roman script letters without
accents can be used with numerals, space, period (.) and forward slash (/).
The screen first shows the present equipment number or a series of blank spaces
if nothing has been set yet. If you wish to change or add to the equipment
number then activate the lower right (+/-) switch otherwise move on by activating
the OK switch.
You change the first character position of the equipment number using the lower
switches. A flashing indicator appears under the character you are changing.
When the first character is correct use the top right ( ) switch to move to the next
character position and so on. You can leave a character blank by moving straight
to the next position.
If you make a mistake you cannot move the cursor backwards but you can start
the whole entry process again (see below). After entering all the characters you
want you can move quickly through any remaining blank spaces by holding the
top right ( ) switch for a few seconds.
When you have moved past the last character position the screen changes and
you have the chance to make more changes by activating the (+/-) switch again.
If the equipment number is now correct you must save the changes you have
made by activating the OK switch.

13.13 MAINS FREQUENCY

Choose 50 or 60 Hz according to the nominal frequency of your local mains
supply. This setting helps the instrument to reject interference.
After you activate the OK switch for this screen you have finished the
configuration cycle.

13.14 LEAVING CONFIGURATION

Remove the key plug if you have made all the changes you need. You can
continue through another configuration cycle if you need to make more changes.
You can remove the key plug at any time or switch the instrument off using the
I/O (on/off) switch but remember that any changes you make on a screen are
only saved when you activate the OK switch for that screen.
The configuration cycle will be interrupted if you connect the instrument to a
battery charger.
Page 40 of 112
HAFFMANS RPU 352, RPU 353, RPC 80 RPC 50
RPU 352, RPU 353, RPC 80, RPC 50